TECHNOPOOL3
Dosing systems for swimming pools
14.0 APPENDIX 4 – HYDROGEN PEROXIDE MEASURE CELL (HP)
Operating temperature: from 1 to 45° C.
Maximum working pressure: 1 bar.
Output signal: 4÷20 mA current proportional to the ppm reading.
Power supply: 10...30 Vdc.
Electrolyte life: from 3 to 6 months.
Membrane life: approx. one year (depends greatly on the quality of the water).
A device must be used for the calibration, in order to determine the content of hydrogen peroxide according to the DPD
method (it is recommended to use a photometric instrument).
The reading of the hydrogen peroxide measure cell supplied is slightly affected by the variations in pH. (range 2÷11
pH).
The recommended flow rate through the flow sensor is 30 L/h. A minimal water flow is required. The measure slightly
depends on the water flow rate.
The measure is compensated in temperature.
The probe must never work with no water.
The probe cannot be used in water with no disinfectant for more than a day.
Attention: accumulated sediment or biological contamination on the membrane can interfere with the
measure.
Before storing the probe, remove the membrane cap and rinse it together with the electrode in clean water. Then dry
all the parts in a dust-free area. Then close the membrane cap without tightening to protect the electrode.
Attention: the membrane must not touch the measure electrode.
When the probe is to be used again, clean the tip of the electrode with the special sand paper and use a new
membrane cap.
Used membrane caps that have been used for at least a day, cannot be stored and re-used.
If calcium forms on the membrane cap, it is recommended to place it in a solution with 1% hydrochloric acid for a few
hours. Then wash it thoroughly with clean water and dry it before using it again.
Description of the hydrogen peroxide electrode
1.
PG7 cable gland.
2.
Cover.
3.
O-Ring.
4.
2-pole terminal for the electrical connection.
5.
Electrode holder.
6.
Counter electrode (stainless steel).
7.
O-Ring.
8.
Reference electrode.
9.
Measure electrode.
10. Transparent cover.
11. Membrane cap.
12. Membrane covering (stainless steel).
13. Membrane.
14. Valve.
Installation procedure
Probe polarization is required on the first installation. After having connected the probe to the Technopool3,
immerse it in a container of water at 0 ppm and leave it for an hour.
The membrane is extremely sensitive! When removing or tightening the membrane cover, make sure the
membrane is not damaged with the high or low pressure.
If calibration is not possible due to the values being unstable or too low, the electrolyte must be changed.
ADSP7000570
Electrical connections
1. Connect the wires to the terminal (4): brown (+) / blue (-).
2. Tighten by hand (2) until the O-Ring (3) isolates the system.
3. Close the PG7 cable gland securely (1).
rev. 1.6
09/01/2014
APPENDIX 4
ENGLISH
66/136