Calibration Of Ph Electrode; Calibration Of The Redox Electrode - aqua Technopool3 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TECHNOPOOL3
Dosing systems for swimming pools
4.10 C
ALIBRATION OF P
The calibration of the pH electrode is carried out on two points, therefore it is requested to have the following
material always available:
Select the option pH Calibration from the installer menu and follow the instructions displayed.
Once carried out the pH 7 calibration, rinse the electrode with water, so to prevent the pH 4 solution from
polluting.
Once the procedure is completed, the quality of the calibrated electrode will be displayed; if the quality of the probe
results 25% or less, carry out a new calibration.
Should the result still be 25% or less, it is recommended to replace the probe.
4.11 C
ALIBRATING THE REDOX ELECTRODE
The calibration of the redox electrode is carried out only on one point, therefore it is requested to have the following
material always available:
Select the option Rx Calibration from the user menu and follow the instructions displayed.
Once the procedure is completed, the quality of the calibrated electrode will be displayed; if the quality of the probe
results 25% or less, carry out a new calibration.
Should the result still be 25% or less, it is recommended to replace the probe.
ADSP7000570
H
ELECTRODE
pH 7 buffer solution.
pH 4,01 buffer solution.
Container with water for cleaning the electrode (i.e. a glass).
IT IS DISPLAYED FOR
475 mV buffer solution.
rev. 1.6
T
/
H/R
H/R
IME
P
X AND P
X TECHNOPOOL
09/01/2014
QUICK PROGRAMMING
ENGLISH
3.
55/136

Publicité

Table des Matières
loading

Table des Matières