4.12 C
ALIBRATING THE CHLORINE ELECTRODE
The calibration of the chlorine electrode is carried out only on one point, therefore the following material must be
readily available:
Select CI Calibration in the user menu and follow the instructions that scroll on the display.
When the system acquires the calibration value, two parameters regarding the measure are displayed:
the current value (0.00÷20.00 mA).
the measured chlorine value (0.00÷20.00 ppm).
Furthermore, the value detected by the DPD1 reference instrument can be associated with the value measured by
the system using the arrow keys in the Cl set menu.
The value that can be set ranges between 0.00 and 20.00 ppm.
4.13 C
ALIBRATING THE HYDROGEN PEROXIDE ELECTRODE
The calibration of the hydrogen peroxide electrode is carried out only on one point, therefore the following material
must be readily available:
Select HP Calibration in the user menu and follow the instructions that scroll on the display.
When the system acquires the calibration value, two parameters regarding the measure are displayed:
the current value (0.00÷20.00 mA).
the measured hydrogen peroxide value (0.0÷500.0 ppm).
Furthermore, the value detected by the DPD4 reference instrument can be associated with the value measured by
the system using the arrow keys in the HP set menu.
The value that can be set ranges between 0.0 and 500.0 ppm.
ADSP7000570
IT IS DISPLAYED FOR TECHNOPOOL
DPD1 reference instrument.
DPD4 reference instrument.
rev. 1.6
3
IT IS DISPLAYED FOR TECHNOPOOL
09/01/2014
QUICK PROGRAMMING
H/C
J
P
L
3
H/HP
P
ENGLISH
56/136