Les langues disponibles

If motors with integrated control units have to be removed / mounted, all cable connections to the other motors, as well as to the
hand control, must be removed / fitted.
All control units have a lock cam that must be loosened before the plugs can be pulled out of the control unit.
1.
Release the lock cam by pressing the small locking pins out with a screwdriver.
2.
When mounting is complete, the lock cam must be fitted again.
3.
Press down on the lock cam so the locking pins engage.
- Mounting the motor
1.
Place the motor in the fork mounts and pull the split pin through.
2.
Lock the split pin.
3.
Insert the plugs into the control unit. Take care that the plug is
the right way round and inserted correctly into the control unit.
4.
Replace the lock cam.
5.
Make sure that all the working parts function correctly.
- Removing the motor
When removing the height adjustable motor, the mattress support must be
held up by two persons.
1.
Remove the lock cam and pull the cables out of the control unit.
2.
Open the split pins at both ends of the motor and pull them out.
3.
The height adjustable motor can now be removed and the mattress support lowered
into the lowest position.
(On some models, the ends of the motor are secured by bolts and these must be removed).