Insulating Resistance (Riso ) - Benning IT 105 Notice D'emploi

Masquer les pouces Voir aussi pour IT 105:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Buzzer (F1):
With the buzzer being enabled, a continuous acoustic signal will be emitted at a measured value of < 1 Ω.
Null balance (F3):
In order to carry out a null balance of the measuring lead resistance, connect the measuring leads with
each other by means of the alligator clips and press the function key F3 3 until the symbol
on the digital display 2 .
Measuring lead resistances can be compensated by up to 10 ohms.
AUTO start (F4):
With the AUTO start function being enabled, the continuity test will be started automatically if the resis-
tance value applied to the measuring probes is < 20 kΩ. The function remains stored even after switching
off the tester.
-
Connect the measuring leads to the BENNING IT 105 as shown in figure 2 and apply them to the test
object.
-
The continuity test is started automatically if the AUTO start function has been enabled via the function
key F4 3 . Alternatively, press and hold the TEST key 4 to start the continuity test.
-
In order to reverse the polarity of the testing current, repeat the measurement with inverted measuring
leads at the test object.
-
The resistance value is shown on the main display K and the testing voltage is shown on the sec-
ondary display J.
8.4 Insulating resistance (R
Measurements of the insulating resistance must be carried out at idle system parts only.
If a voltage of > 30 V AC/DC is applied to the test object, a flashing warning symbol
a pulsating acoustic signal will warn you of an external voltage being applied. The external
voltage is indicated on the digital display 2 and measurement will be blocked. Make sure
that the circuit is free of voltage and repeat the measurement.
-
Select the desired function R
-
The symbols of the function keys F1 A to F4 D are briefly shown on the digital display 2 . You can use
the function keys F1 to F4 3 to make the following settings which will be stored until they are changed
for the next time:
F1
Buzzer (F1):
With the buzzer being enabled, a pulsating acoustic signal will be emitted at a measured value of < 1 MΩ.
Lock (F2):
The lock function enables a continuous measurement of the insulating resistance without having to press
and hold the TEST key 4 again. For continuous measurement, first press the function key F2 3 and then
the TEST key 4 . The digital display 2 shows the LOCK symbol
ously applied to the measuring probes. The lock function can be stopped by pressing the function key F2
3 or the TEST key 4 .
-
Connect the measuring leads to the BENNING IT 105 as shown in figure 2 and apply them to the test
object.
-
Press and hold the TEST key 4 to start an insulating resistance measurement.
-
The resistance value is shown on the main display K and the testing voltage is shown on the sec-
ondary display J.
04/ 2020
)
ISO
(250 V D), 500 V C or 1000 V B) with the rotary switch 1 .
ISO
F2
BENNING IT 105
F3
B and the testing voltage is continu-
C is shown
and
F4
22