Connexion « EtherCAT »
1
Idx
sIdx
Nom
51
Prs.Enable
7000
Turbo Control
01
Turbo Start Stop
02
Turbo Reset Alarm
03
Turbo Low Speed
04
Motor off
FBF0
Device Reset Command
01
command
02
status
03
response
FBF1
Exception Reset Command
01
command
02
status
03
response
FBF2
Store Parameters Command
01
command
02
status
03
response
FBF3
Calculate Checksum Command
01
command
02
status
03
response
1. Idx = Indice, sIdx = Sous-indice, hexadécimal
2. RO = lecture seule, RW = lecture/écriture, P = convient pour les données de processus
1
Idx
sIdx
Nom
4001
Cfg Pump Operation
21
Intf. Configuration
84
Nominal Speed
85
Process Gas
86
Cfg Ready for Process
87
Speed Switchpoint 1
88
Speed Switchpoint 2
89
Enable Run-up Timer
8A
Run-up Time
4002
Cfg Pump Power/Temperature
11
Max. Power
4005
Cfg Component Operation
18
Données de configuration (commande -> TC 400 TC)
2
Type
Accès
USINT
RW P
BOOL
RW P
BOOL
RW P
BOOL
RW P
BOOL
RW P
BYTE()
RW
USINT
RO
BYTE()
RO
BYTE()
RW
USINT
RO
BYTE()
RO
BYTE()
RW
USINT
RO
BYTE()
RO
BYTE()
RW
USINT
RO
BYTE()
RO
2
Type
Accès
USINT
RW
UDINT
RW
USINT
RW P
USINT
RW P
USINT
RW P
USINT
RW P
BOOL
RW P
USINT
RW P
USINT
RW P
Description
Activation du capteur de vide éle-
vé
0 = arrêt
1 = marche
2 = marche, lorsque le point de
commutation de la vitesse de
rotation est atteint
3 = marche, lorsque la valeur
est descendue sous le point de
commutation de la pression
État de la pompe (0 : arrêt, 1 :
marche)
0->1 : acquittement de panne
Veille (0 : arrêt, 1 : marche)
Moteur (0 : marche, 1 : arrêt)
Uniquement lorsque la pompe est
à l'arrêt complet
{74 h, 65 h, 73 h, 65 h, 72 h, ...
...0x00} : réinitialisation standard
...0x66} : réinitialisation usine
{74 h, 65 h, 73 h, 65 h, 72 h}
R
: 81 00 00 00
H
W
: 01 00 00 00
H
Description
0 : relais standard
1 : inversé
-1
min
0 : gaz lourds
1 : gaz légers
2 : hélium
0 : point de commutation 1
1 : point de commutation 1 et 2
% de la vitesse nominale de rota-
tion
% de la vitesse nominale de rota-
tion
min
%