RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE EQUIPMENT OR SYSTEMS - FOR HANDHELD
ECG MONITOR THAT ARE NOT LIFESUPPORTING
Recommended separation distances between portable and mobile RF communica-
tions equipment and the equipment
The equipment is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the equipment can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the equipment as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output
of transmitter
150 kHz to 80 MHz
w
[ 3.5
d =
] √
V1
0.01
-
0.1
-
1
-
10
-
100
-
For transmitters rated at a maximum output power not listed above the recommended sepa-
ration distance d in metres (m) can be estimated using the equation applicable to the fre-
quency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and reflection from structures, objects and people.
155
BewellConnect - BW-HR1 - User Manual - 042016
Separation distance according to
frequency of transmitter M
80 MHz to 800 MHz
800 MHz to 2.5 GHz
P
d =
1.2 P
√
d =
0.12
0.38
1.2
3.8
12
GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY –
FOR HANDHELD ECG MONITOR THAT ARE NOT LIFE- SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of the equipment should assure that it is used in such an environment.
Immunity test
Radiated RF
IEC 61000-4-3
2.3 P
√
0.23
0.73
2.3
7.3
23
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by ab-
sorption and reflection from structures, objects and people.
IEC 60601
Compliance
Electromagnetic environment
test level
level
3 V/m 80
3 V/m
Portable and mobile RF communica-
MHz to 2.5
tions equipment should be used no
GHz
closer to any part of the equipment, in-
cluding cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2 P √ 80 MHz to 800 MHz
d = 2.3 P √ 800 MHz to 2.5 GHz
where p is the maximum output
power rating of the transmitter in watts
(W) according to the transmitter manu-
facturer and d is the recommended sep-
aration distance in metres (m).
Field strengths from fixed RF transmit-
ters, as determined by an electromag-
netic site survey,
the compliance level in each frequency
b
range.
vicinity of equipment marked with the
following symbol:
BewellConnect - BW-HR1 - User Manual - 042016
– guidance
a
should be less than
Interference may occur in the
156