Minebea Intec YPS08-Y Manuel D'installation page 40

Publicité

Les langues disponibles

Les langues disponibles

Class Division / Class Zone:
Hazardous Location
Class I, II, III, Div 1, GP A,B,C,D,E,F,G
Class I, Zone 1, Grp. IIC
Class II, Zone 21, Grp. IIIC
Class III
Minebea Intec
6)
Device
Note:
1)
USA: The installation must be in accordance with the National Electrical Code
504 or 505 and ANSI / ISA-RP 12.6.
Canada: The installation must be in accordance with the Canadian Electrical Code
2)
Mains cable: No connection to any device that uses or generates in excess of 250Vrms or
250Vdc.
3)
USA: The apparatus must be connected to a suitable ground electrode per National Electrical
®
Code
, NFPA 70, Article 504 or 505. The resistance of the ground pad must be less than 1 ohm.
Canada: The apparatus must be connected to a suitable ground electrode per Canadian
®
Electrical Code
4)
Cable of the power supply up to 50m (165 ft).
5)
Option: separable connection
6)
Any intrinsically safe Minebea Intec device suitable for use in the hazardous location classified
above.
7)
WARNING: SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
AVERTISSMENT: LA SUBSTITUTION DE COMPOSANTS PEUT COMPOROMETTRE LA
SECURITE INTRINSEQUE.
8)
No changes to this drawing without prior certification.
2019-02-05
T. Hiller
1)
5)
3)
-10°C .... +40°C (+14°F .... + 104°F)
, Section 18. The resistance of the ground pad must be less than 1 ohm.
Non-Hazardous Location
4)
Uo = 12.6V
Io = 205mA
Po = 2.59W
Co = 1μF*
Lo = 100μH*
*: incl. 50m cable
Control Drawing
67005-751-07
2)
Minebea Intec
Model
YPS08-Z...
3)
®
, NFPA 70, Article
®
, Section 18.
YPS08-Z...
00
1
Sheet
Revision
1
of