17
Technical Information
RS 232/RS 485 Communication Description
N.
202
203
204
205
206
300
301
302
320 to
399
400
402
404
500
503
504
236/252
Read/
Data-
Description
Write
type
R
N
Pump power in W (pump current
x pump voltage duty cycle)
R
N
Driving frequency in Hz
R
N
Pump temperature in °C
R
N
Pump status
R
N
Error code
R
N
Cycle time in minutes (zeroed by
the reset command)
R
N
Cycle number (zeroed by the
reset command)
R
N
Pump life in hours (zeroed by the
reset command)
Reserved to Agilent service
R
A
CRC EPROM (QE)
R
A
CRC Param. (PA)
R
A
CRC Parameter structure
Reserved to Agilent service
R/W
N
RS 485 address
R/W
L
Serial type select
TV 801 Navigator User Manual / 87-900-953-01
Admitted Values
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description:
see the following
figure
0 to 999999
0 to 9999
0 to 999999
QE8XXXX (where
"XXXX" are
variable)
PA8XXXX (where
"XXXX" are
variable)
"XXXX"
0 to 31
(default = 0)
0 = RS 232
1 = RS 485
(default = 0)
(G)