Technical Information
RS 232/RS 485 Communication Description
Window-Meanings
Tab. 7
N.
000
008
100
101
102
103
104
105
106
107
234/252
Window-Meanings
Read/
Data-
Description
Write
type
R/W
L
Start/Stop (in remote mode the
window is a read only)
R/W
L
Remote (default) or Serial
configuration
R/W
L
Soft Start
(write only in Stop condition)
R/W
N
Set Point type
R/W
N
Set Point threshold (expressed in
Hz, mA or s)
R/W
N
Set Point delay: time between
the pump start and the set point
check (seconds)
R/W
L
Set Point signal activation type:
the signal can be "high level
active" or "low level active"
R/W
N
Set point hysteresis (in % of
threshold)
R/W
L
Water cooling
R/W
L
Active Stop (write only in stop)
TV 801 Navigator User Manual / 87-900-953-01
Admitted Values
Start = 1
Stop = 0
Remote = 1
Serial = 0
(default = 1)
YES = 1
NO = 0
0 = Frequency
1 = Current
2 = Time
(default = 0)
(default = 867)
0 to 99999
(default = 0)
0 = high level
active
1 = low level
active
(default = 0)
0 to 100
(default = 2)
0 = NO
1 = YES
0 = NO
1 = YES
(G)

Publicité

loading

Ce manuel est également adapté pour:

Sq 336Sq 306Sq 337