Distriwel –August 2017
Servicing and maintenance
Summary table
Component
Device
Piston
UV lamp
Disposable plastic cup
Dispensing hose
Silicone connection
Fixed plate
Sliding plate
Regular checks
If the agar has run onto the cross, it is vital to clean the cross after the cycle and to dismantle it
Cleaning
Of the cross and disk
It is recommended to clean the device fully on a daily basis by dusting it and washing with hot water and
alcohol at 70°C.
You will have to remove the carousel and cross using the following procedure:
8-
Switch the device off.
9-
Remove the altuglas lid
10-
Remove the carousel's tightening handle.
11-
Remove the carousel from its shaft, by pulling it upwards. The easiest way to do this is by holding it with
one hand at the bottom and the other hand at the top.
12-
Next, remove the cross's tightening handle (Fig. 1), and then the cross itself.
13-
Once you have removed the cross, remove the plate (Fig. 3 )
14-
Wash the device and all its components using hot water and alcohol at 70°C.
Operation
After each
Cycle
Dismantling and
Cleaning
Cleaning
Replacement
Replacement
Cleaning /
Autoclaving
Replacement
Replacement
Lubrication
Lubrication
Frequency
Daily
C
C
C
C
Variable
400 hours
Each week
100 hours
Every 2 months
Twice / month