Distriwel –August 2017
Make sure to always have several sets of hoses available. Three hoses are delivered with the device (ref.:).
After each cycle, rinse the hoses immediately in hot water, to prevent the agar blocking the hose.
Lubrication
Fixed plate
To make it easier to slide the dishes onto the fixed plate, lubricate the plate with a liquid silicone spray (type KF 6102)
every 2 months.
Sliding plate
If using Petri dishes that leave a white deposit - in addition to the general daily cleaning, you will have to lubricate the
sliding plate at least twice a month
Maintenance
UV lamp (ref.: )
The UV lamp must be changed after every 400 hours of use.
•
Switch the device off.
•
Open the altuglas lid.
•
Turn the lamp a quarter turn and pull it upwards to remove.
•
Insert the replacement lamp by repeating this action in reverse.
Silicone connection
The silicone connection, located on the pump rotor during the dispensing operation, must be changed after every 100
hours of use in order to guarantee a regular volume and flow rate.
Only use the hose supplied by Alliance Bio Expertise Ref. ?