Calibrate the pump as seen for pump N°1 (previous chapter), and press the green button.
Press the green button to continue without calibration, or press the arrow to calibrate pump n°2. Use a
measuring cylinder or a scale to compare the dispensed to the set up volume.
Enter the measured volume, and do it again if necessary.
Once correct, press the green button to validate, the following screen appears.
Press the green button to start the selected program.