Servicing and maintenance

Summary table

Element
Bactystopper
MEDW3003
Sampling tube O ring
(X5)
MEDW4101
O ring for addition cap
MEDW4103
Lid seal
MEDW4102
Water
inlet
filter
(Option)
MEDW3008
Stirrer bearing kit
MEDW4100
Ring seal on stirrer axis
MEDW4008

Temperature probe calibration

the manufacturer strongly recommends calibrating the temperature probe once a
year.
Fill the vessel of 10 liters of water
Close the lid
Remove the sampling cap from the lid
Insert reference temperature probe (42 cm long probe with specific
adapter)
Launch a cycle with a program at 121°C of 15 min speed 4 for the first
step and 42°C speed 7 for the second step
Read the displayed temperature of the MEDIAWEL during the first step
(after 5 min of stability at 121°C) and read the one of the reference
thermometer at the same time and write them down
Read the displayed temperature of the MEDIAWEL during the second step (after 5 min of stability
at 42°C) and read the one of the reference thermometer at the same time and write them down
Stop the cycle
Fréquency
Operation
1 cycle
Replacement
Replacement
Replacement
Replacement
Cleaning
Replacement
Replacement
20 cycles
2 months
Regular
check
required
especially
if flamed
84
6 months
THERMOMETER WITH PROBE
ON REQUEST
year