How do I evaluate my blood pressure?
Table for classifying blood pressure values in adults in accordance
with the World Health Organisation (WHO) in 2003. Data in mmHg.
Range
Systolic Diastolic Recommendation
blood pressure too low  100
1. blood pressure
100 - 120 60 - 80
optimum
2. blood pressure normal 120 - 130 80 - 85
3. blood pressure slightly
130 - 140 85 - 90
high
4. blood pressure too
140 - 160 90 - 100 Seek medical
high
5. blood pressure far too
160 - 180 100 - 110 Seek medical
high
6. blood pressure
180 
dangerously high
The higher value is the one that determines the evaluation.
Example: a readout value between 150/85 or 120/98 mmHg
indicates «blood pressure too high».
2. Using the Device for the First Time
Inserting the batteries
After you have unpacked your device, first insert the batteries. The
battery compartment 5 is at the back of the device. Insert the
batteries (4 x 1.5 V, size AA), thereby observing the indicated
polarity.
Selecting the correct cuff
Microlife offers different cuff sizes. Select the cuff size to match the
circumference of your upper arms (measured by close fitting in the
centre of the upper arm).
Cuff size
for circumference of upper arm
S
17 - 22 cm (6.75 - 8.75 inches)
M
22 - 32 cm (8.75 - 12.5 inches)
L
32 - 42 cm (12.5 - 16.5 inches)
M - L
22 - 42 cm (8.75 - 16.5 inches)
Optional preformed cuffs «Easy» are available.
Only use Microlife cuffs.
Contact your local Microlife Service if the enclosed cuff 6 does
not fit.
BP 3AG1
All manuals and user guides at all-guides.com
 60
Consult your
doctor
Self-check
Self-check
Consult your
doctor
advice
advice
110 
Urgently seek
medical advice!
Connect the cuff to the device by inserting the cuff connector 7
into the cuff socket 3 as far as it will go.
3. Taking a Blood Pressure Measurement using this
Device
Checklist for taking a reliable measurement
1. Avoid activity, eating or smoking immediately before the
measurement.
2. Sit down for at least 5 minutes before the measurement and relax.
3. Always measure on the same arm (normally left).
4. Remove close-fitting garments from the upper arm. To avoid
constriction, shirt sleeves should not be rolled up - they do not
interfere with the cuff if they are laid flat.
5. Always ensure that the correct cuff size is used (marking on the
cuff).
• Fit the cuff closely, but not too tight.
• Make sure that the cuff is positioned 2 cm above the elbow.
• The artery mark on the cuff (ca. 3 cm long bar) must lie over
the artery which runs down the inner side of the arm.
• Support your arm so it is relaxed.
• Ensure that the cuff is at the same height as your heart.
6. Press the ON/OFF button 1 to start the measurement.
7. The cuff will now pump up automatically. Relax, do not move
and do not tense your arm muscles until the measurement
result is displayed. Breathe normally and do not talk.
8. When the correct pressure is reached, the pumping stops and the
pressure falls gradually. If the required pressure was not reached,
the device will automatically pump some more air into the cuff.
9. During the measurement, the heart symbol AM flashes in the
display and a beep sounds every time a heartbeat is detected.
10.The result, comprising the systolic 9 and the diastolic AT blood
pressure and the pulse rate AK is displayed and a long beep is
heard. Note also the explanations on further display symbols in
this booklet.
11.When the device has finished measuring, remove the cuff.
12.Switch off the device. (The monitor does switch off automatically
after approx. 1 min.).
You can stop the measurement at any time by pressing the
ON/OFF button (e.g. if you feel uneasy or an unpleasant
pressure sensation).