Sirona Dental Systems GmbH
Operating Instructions PerioScan
13
Electromagnetic compatibility
13.1
Electromagnetic emission
Emission measurement
RF emissions according to
CISPR 11
RF emissions according to
CISPR 11
Harmonics
according to IEC 61000-3-2
Voltage fluctuations/Flicker
according to IEC 61000-3-3
61 29 626 D 3496
D 3496.201.02.05.02
All manuals and user guides at all-guides.com
Conformity
Electromagnetic environment guidelines
Group 1
The UNIT uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class B
The UNIT is intended for use in all facilities, including residen-
tial areas and in any facilities connected directly to a public
power supply providing electricity to buildings used for resi-
not applicable
dential purposes.
Power output
< 50 W
not applicable
no significant
flicker
13 Electromagnetic compatibility
The UNIT is intended for operation in the electromag-
netic environment specified below.
The customer or user of the UNIT should make sure that
it is used in such an environment.
The unit meets the requirements for electromagnetic
compatibility (EMC) in compliance with IEC 60601-1-2:
2001.
37