Dynamic Weighing - Steinberg Systems SBS-KW-1000/10O Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 32
SCALE CONFIGURATION
ENTERING THE MENU
Press
twice, to enter Password mode.
The Scale displays P0000. Input Scale Configuration password P0258 . To do so, press
and
to modify the digits. To move the digits, press
Press
to confirm the input and enter the Scale Configuration.
Scale displays SCALE.
DISPLAY RESOLUTION
The display resolution can be set to: The display resolution can be set to: 0.001, 0.002, 0.005, 0.01, 0.02, 0.05,
0.1, 0.2, 0.5, 1, 2, 5, 10, 20, or 50
Press
to enter the Display Resolution. E--- shows.
Press
or
and
to change the resolution value.
AUTO-ZERO RANGE
EN
The Auto-Zero range can be set to: 0% (disabled), 2 (±2%FS), 3 (± 3%FS), 4 (±4%FS), 10 (±10%FS), 20
(±20%FS), 50 (±50%FS), 100 (±100%FS). It is set to ±20%FS by default.
Press
to enter the Auto-Zero range setting.The scale displays
Press
or
and
, to change the value.
MANUAL ZERO RANGE
The manual zero range can be set to: 0% (disabled), 2 (±2%FS), 3 (±3%FS), 4 (±4%FS), 10 (±10%FS), 20
(±20%FS), 50 (±50%FS), 100 (±100%FS).
Press
to enter the Manual-Zero range setting.
The scale displays the Manual-Zero range,
is within the Manual-Zero range.
Press
or
and
, to change the value.
ZERO TRACKING RANGE
The Zero-Tracking range can be set to: 0 (disabled), 0.5 (±0.5e), 1.0 (±1.0e), 1.5 (±1.5e), 2.0 (±2.0e), 2.5
(±2.5e), 3.0 (±3.0e), 3.5 (±3.5e), 4.0 (±4.0e), 4.5 (±4.5e), 5.0 (± 5.0e).
Enabling the Zero-Tracking will enhance the scale tempe¬rature and drift the performance!
REV. V 2016
or
or
.
.
indicating 4% F.S. Zero is only allowed when the weight
26
Press
, to enter the Zero tracking range setting. The scale displays the Zero-Tracking range
The scale displays active value, e.g. 0.5 for the 0.5e value.
Press
or
and
, to change the value.
ZERO RANGE
Zero Range defines the range that scale must fall into before activating the accu¬mulation operation. When
the load is removed from the scale, the weight left must be lighter than the value set.
Press
, to enter the Zero range setting.
Press
or
and
, to change the value. Press
Input the Zero Range value. Press
, to set the decimal point. Zero Range can be set from 0e to 3000e.
It is set to „5e" by deafult.
ZERO-SAVING
When Zero-saving is enabled, Auto Zero is disabled automatically. Scale calcu¬lates weight based on the last
Zeroing action. It is disabled by default.
Press
, to enter Zero Saving.
shows.
Press
or
and
, to change the status to „on" (enabled) or „off" (disabled).
ANTI-MOTION
At the cost of measuring time, the Anti-Motion function intelligently settles the determination of weight
when the scale is in motion. The weaker Anti-Motion is, the faster the scale displays the weight, but the longer
it takes to be stable.
Anti-Motion can be set to: 0 (off), 1 (weakest), 2 (weak), 3 (normal), 4 (strong), 5 (strongest).
Press
, to enter the Anti-motion setting.
Press
or
and
, to change the value.

DYNAMIC WEIGHING

In some special application where the scale´s accuracy is not so important as the scale´s stability for the
weight reading, the dynamic weighing can be a useful function to set the weight reading faster.
Press
, to enter the Dynamic Weighing setting.
Press
or
and
, to change the status to „on" (enabled) or „off" (disabled).
REV. V 2016
shows.
or
, to move the digit.
shows
shows.
27
EN