SPECIFICATIONS
PD1000A Series Conserving Device Specifications
Power Supply .......................................................................................................................................................................................................................... (2) Standard "AA" alkaline or NiMH batteries.
Operational Voltage Range .................................................................................................................................................................................................................................................... 2.3 to 3.6V DC
Power Requirements ..................................................................................................................Average steady state "ON" current 1.6 uA. Batteries other than alkaline or NiMH are not recommended
due to the capacity needed for operation and battery life of the unit. Typical new battery life is 200 hours when used at 25°C, 2 LPM and 20 BPM. Settings and breath rate will affect battery life.
After the Low Battery (flashing red) light illuminates, the unit will continue to operate about four hours when used at 25°C, 20 BPM and the 6 LPM setting. Settings, breath rate, and battery condi-
tions will affect use times. Refer to local regulations for battery recycling and/or disposal requirements.
Degree of Protection Against Electric Shock ............................................................................................................................................................................................................... TYPE BF applied part
Modes of Operation .........................................................................................................................................................................................................................................................Continuous / Pulsed
Operating Temperature Range ........................................................................................................................................................................................................................... 14° to 104°F (-10° to 40°C)
Operating Pressure Range ...............................................................................................................................................................................................500 to 2250 PSIG (34 to 155 bar—tank pressure)
Operating Atmospheric Conditions ....................................................................................................................................................................................................................................... 500 to 1020 hPa
Operating Humidity Range ............................................................................................................................................................................................................................0 to 95% R. H. non-condensing
Storage and Transportation Temperature Range (Tested at ~933 hPa) .......................................................................................................................................................... .-40° to 158°F (-40° to 70°C)
Storage and Transportation Humidity Range (Tested at ~933 hPa) .............................................................................................................................................................0 to 95% R. H. non-condensing
Expected Shelf and Service Life (excluding batteries) ...................................................................................................................................................5 years based on 4 hours use per day at 20 BPM
Degree of Protection Against Ingress of Liquids ..................................................................................................................................................................................................................................... IPX1
Safety Standard .................................................................................................................................................................................................................................................... meets ISO 18779: 2005(E)
Approval Body And Safety Standard .................................................................................................................. Approved by CSA to: IEC 601-1; CAN/CSA-C22.2 No 601.1-M90 as ordinary equipment
US Patents ..................................................................................................................................................................................................................................................4,519,387; 5,755,224; 4,457,303
®
Certified to CAN/CSA C22.2 No. 601.1-M90
iFill Oxygen Cylinder w/Integrated PulseDose Regulator "Europe (Pi-mark)"
Product # "Europe (Pi-mark)"
PD1000A-I-ML6
PD1000A-I-C
PD1000A-I-D
PD1000A-I-E
iFill Oxygen Cylinder w/Continuous Flow Regulator "Europe (Pi-mark)"
Product # "Europe (Pi-mark)"
535I-ML6-CF
535I-C-CF
535I-D-CF
535I-E-CF
All iFill Oxygen Cylinders Specifications
Operating Temperature Range. ............................................................................................................................................................................................................................. 41° to 104°F (5° to 40°C)
Operating Pressure Range ................................................................................................................................................................................................. 500 to 2250 PSIG (34 to 155 bar) tank pressure
Operating Atmospheric Conditions ....................................................................................................................................................................................................................................... 500 to 1020 hPa
Operating Humidity Range ............................................................................................................................................................................................................................0 to 95% R.H., non-condensing
Storage and Transportation Temperature Range ................................................................................................................................................................................................ -4° to 130°F (-20° to 54°C)
Storage and Transportation Humidity Range ............................................................................................................................................................................................. Up to 95% R.H., non-condensing
Degree of protection against ingress of liquids .....................................................................................................................................................................................................................................NONE
Safety Standard ............................................................................................................................................................................................................................................meets TPE Directive 99/36/ECπ
Specifications subject to change without notice.
NOTE- Degradation of performance may occur if unit is operated outside of specified operating parameters.
ACCESSORIES
The accessories below are approved for use with the DeVilbiss unit:
Carry Bags
C Cylinder bag ....................................................................................................................................................................................................................................................................EX3000D-651
D Cylinder bag ....................................................................................................................................................................................................................................................................EX3000D-652
ML6 Cylinder bag ................................................................................................................................................................................................................................................................EX3000D-654
Cylinder cart (E Cylinder) ..................................................................................................................................................................................................................................................................... CT001
There are many types of oxygen tubing and cannulas that can be used with this device. Certain accessories may impair the device's performance. Use only standard nasal cannula capable of
supporting a minimum flow rate of 10 LPM with PulseDose delivery. Do not use pediatric (low-flow) nasal cannula or mask with PulseDose delivery. A mask or any nasal cannula can be used with
continuous flow delivery and may be sized according to your prescription as recommended by your homecare provider who should also give you advice on the proper usage, maintenance, and
cleaning.
EN - 8
Weight
4.8 lbs (2.18 kg)
4.38"D x 11.68"L (111mm x 297mm)
5.6 lbs (2.54 kg)
4.38"D x 14.88"L (111mm x 378mm)
7.2 lbs (3.27 kg)
4.38"D x 20.51"L (111mm x 521mm)
9.8 lbs (4.45 kg)
4.38"D x 29.63"L (111mm x 753mm)
Weight
4.6 lbs (2.09 kg)
4.38"D x 11.0"L (111mm x 279mm)
5.4 lbs (2.45 kg)
4.38"D x 14.2"L (111mm x 361mm)
7.0 lbs (3.18 kg)
4.38"D x 19.82"L (111mm x 503mm)
9.6 lbs (4.35 kg)
4.38"D x 28.94"L (111mm x 735mm)
0044
Dimensions
Dimensions
SE-535-CYL