INSTRUCTIONS FOR USE
10. Perform a final check to ensure that the patient's heels are properly seated in
the heels of the Boots and that there are no pressure points on the calf. The leg
must be centered in the Boot to eliminate pressure on the peroneal nerve.
Secure the pad. Over tightening the Velcro strap serves no clinical benefit and
can cause damage to the strap/pad. Gently secure the strap to the boot, never
pulling it at an angle.
11. Grasp and twist Release Handles and simultaneously move legs slowly to desired
height, maintaining leg symmetry.
12. Once lithotomy level has been achieved, check for pressure points and, if
necessary, make appropriate adjustments.
NOTE: Patient leg position symmetry can be documented using leg length
indicator markings on the Stirrup posts.
3.3 Device controls and indicators:
The indicators below are placed on the sole of each boot.
3.4 Storage, Handling and Removal Instructions:
3.4.1
Storage and Handling:
The product should be stored in a clean and safe environment to prevent product
damage. See storage Specifications under Product Specification section.
3.4.2
Removal Instruction:
a. Lower Stirrups by grasping and twisting the Release Handles and simultaneously
down staging the legs slowly to the desired level. A.O.R.N. recommends the
lowering and staging process take two full minutes. Gently remove patient's legs
from Stirrups.
b. Stirrups should be in horizontal position for removal from Rail Clamp slot. Loosen
clamp and remove Stirrups.
Document Number: 80028291
Version: B
This symbol represent the patient's right foot.
This symbol Represent the Patient's left foot
Page 14
Issue Date: 10 AUG 2021
Ref Blank Template: 80025118 Ver. E