TOOLS REQUIRED FOR MOUNTING AND MAINTENANCE OF ALL THE
SOCKETS :
Multifunction key K 00 70 91.
9.1.2. LEAKAGE AND FUNCTIONALITY TEST
•
Test connectivity of counterparts of all installed units, and in the case of
variants with positioning of the quick connector counterpart check the
units for correct orientation.
•
Test the general tightness of the unit with connected as well as discon-
nected counterparts.
•
Clean the unit from possible impurities and check all labels and mark-
ings for integrity and presence.
Some repairs within the framework of the maintenance concerning re-
placement of damaged or missing parts can be carried out by the product
owner. It is possible to replace the following parts only:
•
labels
Use only original spare parts.
All sealing and o-rings must be stored by provider or operator in a dry,
dark and clean storage during the complete lifetime.
9.2. REPAIR
Service, repair and spare parts exchange can only be done by a person
that has the appropriate certifi cates according to the required national
standards and laws for mounting and repairs of concerned gas medical de-
vices. Standards valid for such installations and testing are EN ISO 7396-1
for medical central gas systems and EN ISO 5359 for installations with low
pressure hoses.
For more information and actual list of spare parts contact GCE or distribu-
tor of GCE products.
Use only GCE original spare parts.
10. PRODUCT LIFETIME
10.1. MANUFACTURING DATE AND NUMBER
Number perforated on the housing and quick-coupler is combination of
the following data:
XXXXXXX: batch number
YY/MM: date code
Ex.: Number 7366506 indicates a component with batch number 7366506.
Ex. Number 13/04 indicates a component with manufacturing date - April,
year 2013.
15/84