6.2. CHECKS AFTER INSTALLATION

6.2.1. FUNCTIONAL AND LEAKAGE TEST AFTER INSTALLATION
Check correct functionality of TU by connecting the
probe (connector).
Check if the GCE logo on the button is in the position
as on the picture.
Check the installed TU for leakage. Max. testing pres-
EN
sure must not exceed 24 bar.
Installation and testing procedures must be in compliance with
EN ISO 7396-1 and EN ISO 5359 standards.
7. OPERATION

7.1. BEFORE OPERATION

PREPARATION FOR USE OF TU
Check if the TU (including label) does not show any visible external dam-
age. If there are signs of external damage, ensure the unit is not used
and label it as faulty. Contact the responsible personnel and call service
- see chapter 9.2.
Check visually if the product is clean, if cleaning is needed follow proce-
dures described further in this instruction for use.
TU can be used only for the gas specifi ed on the sleeve. Never try to use
it for di erent type of gas.
7.2. CONNECTING DEVICE TO TU
Any devices used with TU must be compatible with the gas for which TU
is intended for.
Always make sure the device is connected fi rst and then used on the
patient.
7.2.1. CONNECTING DEVICE TO TU
Push the probe (L1) into the TU and lock it in by
turning into working position. Gas fl ow is on the
outlet.
12/84
2
L1
1