6.1.3. QUICK COUPLER INSTALLATION QC
EN
A
Fit a quick coupler (D) into the housing (G)
and tighten by means of pipe wrench (tight-
ening torque 12 Nm).
Fix a cover (recessed C2 or exposed C1) on
the installation box (I).
CE label is located in right bottom corner.
If you have a variant with a lid, assembly the
lid to the sleeve (B1) and fi x by means of the
pin (K).
Fix a sleeve (B) to the cover (C).
Snap the gas specifi c plugs (A) into the
sleeve (B) according to the connected gas
type. If the gas is indicated by two di erent colours keep the order of
colours from left to right according to EN ISO 9170-1 or national standard,
ex. medical air (AIR) - black-white.
B1
C2
C1
B2
D
Make sure the
8/84
G
B2
K
I