For further information about training programs, please contact GCE.
5. PRODUCT DESCRIPTION
MEDICONNECT-DUOLINE consist of three main parts:
A - INLET CONNECTION
Inlet connection serves as inlet for the medical gas.
B - BODY
Body - connects inlet and outlet connection, may also consist a hose or
fixing device.
EN
C - OUTLET KIT
Outlet kit consists of two connections. These connections are designed to
prevent connection of different medical gas than what they are intended
for.
A
* Pictures are only informative, real product might be slightly different.
Exact technical specification is stated in Appendix Nr. 1 - Technical and
performance data.
6. OPERATION
6.1. BEFORE USE
VISUAL CHECK PRIOR TO THE INSTALLATION
•
Check the product for obvious damage (including labels and signs). If
there are marks of external damage, do not use the product and report
its status.
•
Visually check if product is clean. In case of pollution, clean it following
the procedure described in this instruction for use.
•
Using the marking system of GCE or the owner, check if lifetime or date
for disposal of the product has not been exceeded. If the date has been
exceeded, withdraw the product from the operation and report its status.
B
C
A
C
B