H LOCKING PLATE
Locking plate secures the housing with quick coupler in the installation
box.
I INSTALLATION BOX
Installation box is used for installation of TU in or on the wall or into the
pendant channel.
J PLUG
The plug prevents contamination of the installation box.
N FIXING NUT
Nut for fi xation of the plastic covers (B,C) in case of installation without use
of installation box (I).
5.2. TECHNICAL PARAMETERS
DIMENSIONS:
•
Weight: 0,35 kg
•
Height: 65 mm
•
Diameter of installation box: 60 mm
INLET PRESSURE:
•
-0,4 to -0,9 bar for vacuum
•
4 to 5 bar for compressed medical gases
•
7 to 10 bar for air or nitrogen to be used for running the surgical tools
NOMINAL FLOW:
•
40 l/min for compressed medical gases
•
350 l/min for air or nitrogen to be used for running the surgical tools
•
25 l/min for vacuum systems
6. INSTALLATION OF TU, CHECK AFTER INSTALLA
TION
6.1. INSTALLATION
Installation can only be done by a person that has the appropriate certifi -
cates according to the required national standards and laws for mount-
ing and repairs of concerned gas medical devices. Standards valid for
such installations and testing are EN ISO 7396-1 for medical central gas
systems and EN ISO 5359 for installations with low pressure hoses. For
more information contact GCE or distributor of GCE products.
For more information contact GCE or distributor of GCE products.
Do not connect the housing to pipeline with di erent type of gas.
During mounting use always appropriate and functional tools and keep
all safety requirements for operation, transport, storage (see chapter 3).
Ensure that all devices, equipment and also hands and work wear are
kept clean and free from oil and grease.
5/84
EN