6.2 Weighing
7. HYGIENE TREATMENT
38 •
CAUTION!
Incorrect measurement due to force shunt
Operation of the stadiometer affects the weight
display of the scale. Weight values which are
displayed during a height measurement do not
correspond to the patient's actual weight.
Ensure that you do not touch the scale whilst
weight is being measured.
Ensure that the patient does not touch either
the stadiometer or the column of the scale
whilst weight is being measured.
Only read off weight values before or after
measuring height.
The scale functions of the
cal to those of the
seca 704
mation about this in the instructions for use supplied for
the
model.
seca 704
The scale functions of the
cal to those of the
seca 703
mation about this in the instructions for use supplied for
the
model.
seca 703
WARNING!
Electric shock
Set up the device so that the power supply
socket is easy to reach and the device can be
disconnected from the power supply quickly.
Ensure that your local power supply matches
the information on the power supply unit.
Do not touch the power supply unit with wet
hands.
Do not use extension cables or power strips.
Make sure that cables are not pinched or da-
maged by sharp edges.
Make sure that cables do not come into con-
tact with hot objects.
Do not operate the device at an altitude of
more than 3000 m above sea level.
model are identi-
seca 704s
model. You can find infor-
model are identi-
seca 703s
model. You can find infor-