Range of application / Construction and function
Construction of
the sensor The PER sensor consists of 3 main components, the upper section,
28
4

Range of application

IMPORTANT
• Non-compliance with operating conditions specified in
the technical data (see section 14) could lead to a measur-
ing error and to a dangerous over metering within a closed
loop.
• The sensor is not suitable for checking for the absence of
H
O
.
2
2
The DULCOTEST
amperometric sensor for the online determination of the
concentration of hydrogen peroxide. Hydrogen peroxide serves
a.o. as bio-degradable disinfection and oxidation agent. The
sensor can a.o. be used in the following applications:
• Oxidation of drinking water, landfill leachate, contaminated ground
water
• Disinfection of cooling, service, and production water in the
pharmaceutical, food and beverages industry as well as in
swimming pools
• Deodorization (gas scrubber) in municipal and industrial purifica-
tion plants
• Dechlorination in chemical processes
• Chemical bleaching in the timber, paper, textile and mineral
materials industry
• Organic synthesis in the chemical, pharmaceutical, and cosmetics
industry
Please note that for regulation tasks, the response time t
5

Construction and function

the electrode shaft and the diaphragm cap (see fig. 1). The
electrolyte-filled diaphragm cap represents the measuring chamber
into which the measuring electrodes are immersed.
The measuring medium is sealed by a diaphragm in the measuring
chamber.
The electronic amplifier is embedded in a synthetic compound in
the upper section of the shaft.
The sensor has a passive 4-20 mA two-wire interface. Power is
supplied externally from a measuring and control system, e.g.
DULCOMETER
®
D1C for the measured variable H
®
sensor PER 1 is a diaphragm-covered
is 8 min.!
90
O
.
2
2
ProMinent
®