Storage and transport / Application
3
IMPORTANT
• Observe the specified conditions for storage in order to avoid damage
Storage
• Storage period of the sensor including
• Storage period of the electrolyte in
• Storage and transport temperature
• Humidity
Transport
The sensor should be transported only in its original packaging.
4
IMPORTANT
• Ignoring the specified working conditions (see Technical data, Chap-
• The sensor is not suitable for checking the absence of chlorite.
• The presence of reducing reagents may interfere with the measure-
The chlorite measuring cell CLT 1 is a membrane-covered amperometric two
electrode probe. Its main application is in the continuous monitoring of chlorite
concentrations, which can emerge as a by-product of water disinfection using
chlorine dioxide, as high chlorite concentrations can be harmful to health.
Waterworks, drinks manufacturers and other areas of water treatment
represent the typical users of chlorine dioxide in the disinfection of raw water.
By using the DULCOMETER
together with one of the chlorine dioxide systems from ProMinent, the requisite
chlorine dioxide concentration can be regulated in compliance with chlorite
limit values.
28

Storage and transport

or malfunction.
membrane, in the original packaging
the original bottle

Application

ter 14) may lead to incorrect measurements and to dangerous over-
dosing within a control system.
ment.
max. 2 years
max. 2 years
+5 to +50 °C
max. 90 % rel. humidity,
below dewpoint
®
D1C chlorite measuring and control equipment
®