20 Electro-Magnetic Compatibility
• ELECTROMAGNETIC INTERFERENCE WARNING: Some electrically powered equipment
gives off electromagnetic waves which could interfere with the TransAeris stimulator.
When using the TransAeris stimulator around electrical equipment, check the TransAeris
screen to make sure the device is working.
• Do follow the EMC information provided. The TransAeris needs special precautions
regarding electromagnetic compatibility (EMC). To reduce the possibility of interference
on the TransAeris from other electrical equipment or the TransAeris effecting other
electrical equipment, do NOT use cables or accessories with your TransAeris other
than those specified.
• RF COMMUNICATION WARNING: Portable and mobile RF communication equipment
can effect medical electrical equipment.
• The TransAeris should not be used adjacent to or stacked with other equipment and
that if adjacent or stacked use is necessary, the TransAeris should be observed to
verify normal operation in the configuration in which it is used.
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The TransAeris stimulator is intended for use in the electromagnetic environment specified
below. The customer or the user of the TransAeris stimulator should assure that it is used
in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Compliance
Group 1
Class B
Not Applicable
Not Applicable
Electromagnetic Environment – Guidance
The TransAeris stimulator uses RF energy
only for its internal function. Therefore, its
RF emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The TransAeris stimulator is suitable for use
in all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
81 39