The Regulatory Marking of Conformity symbol indicates that the device
C
meets Medical Device Directive 93/42/EEC. This has been certified by
2797
notified body number 2797.
The European Community Representative symbol indicates the
P
identification of the authorized representative for the distribution
of devices into the European community.
The Type BF Applied Part symbol appears on powered equipment that
connects directly to a patient. It is an indication of the degree of protection
provided against electric shock, patient leakage current and patient
auxiliary current.
The On/Off symbol on powered equipment indicates push-button ON/OFF
power control of the device.
The Consult Accompanying Documents symbol appears on powered
equipment indicating that instructions for use must be consulted for safety.
The Ingress Protection (IP) Classification symbol appears on powered
IP
55
equipment indicating that the device is protected from water jets.
Non-ionizing electromagnetic radiation.
MR Unsafe. A device that is known to pose hazards in all MR environments.
MR
Notice of proper disposal.
During NORMAL operation, the check mark will be displayed on the
corresponding stimulating channels indicating stimulus delivered.
During NORMAL operation, the exclamation mark will be displayed on
the corresponding channels requiring attention to check lead connections.
The HOME button used in conjunction with the Unlock button will temporarily
stop stimulation and return the stimulator to Self-test mode.
The UNLOCK button is an interlock button that requires to be pressed
and held for any other function button to operate.
The INCREASE button used in conjunction with the Unlock button allows
the user to increase the selected parameter.
The DECREASE button used in conjunction with the Unlock button
allows the user to decrease the selected parameter.
The channel pointer is displayed below the channels indicating that
the channel has been selected for modification.
ESD sensitive device.
o