10 Perform Cardiac Interaction Test
10.1 PRIOR TO START OF CONDITIONING, CARDIAC INTERACTION
TESTING MUST BE PERFORMED.
• Turn all channels OFF
• Increase Stimulus Intensity to 100%
• Increase Stimulus Frequency to 20Hz
• Burst Mode must be OFF
Test each electrode individually. To do this, turn channel one on. If cardiac interaction is
noted, reduce Stimulus Frequency to 14 and retest. If still noted, reduce percentage of
intensity of the channel until interaction is eliminated. If not eliminated, turn channel off
and remove electrode.
• Repeat for above steps for each channel.
• Repeat test with all channels on.
11 Errors During Normal USE
11.1 ERROR DISPLAYS
Low Battery - This error displays a low battery error during operation.
Electrode Error - This error displays the TransLoc electrode is in need of attention during
operation. The continuity is broken in one or more TransLoc electrodes or the surface
electrode is detached.
74 32
Figure 26 – TransAeris LOW BATTERY Error
Figure 27 – TransAeris TransLoc Electrode Error