14.2 Safety Checks
Perform safety checks on the TransAeris if the device has been dropped or damage
is suspected.
14.2.1 Visual Inspection
Perform the following visual inspections each time the TransAeris is used:
• Check that there is no mechanical or physical damage to the device.
• Confirm that the device will power up and all controls buttons function.
14.2.2 Functional Inspection
Perform the following functional inspections each time the TransAeris is used:
• Verify that the device passes the self-test at power-up.
• Verify that the control buttons and displays function and work properly.
• Inspect all connections and cables. Verify that the patient cables properly connect
to the FrictionLoc connectors and are not damaged.
Caution: Do not attempt to open the TransAeris. The TransAeris is sealed at the factory
and opening the TransAeris will damage the device and void the warranty.
14.3 Service
The TransAeris stimulator is non-serviceable. If the device fails prematurely, contact Synapse
Biomedical Customer Service to return for evaluation.
A serial number identifying each individual TransAeris is printed on the back label of the
device. Reference the device serial number in any correspondence.
15 Product Warranty
The shelf life of the TransAeris is one (1) year and the unconditional product warranty is 30
days of consecutive calendar use. If the TransAeris is within the 1 year shelf life and fails in
less than 30 days of consecutive calendar use, contact the contact Synapse Biomedical
Customer Service for warranty questions.