19 Specifications

19.1 DEVICE SPECIFICATIONS

Table 1. TransAeris – Temporary powered muscle stimulator specifications
Stimulator Specifications
Device Name
Model
Manufacturer
Power Source(s)
Number of batteries
Battery type
Battery size
Operating Temperature
Storage/Transport Temperature
Relative Humidity
Atmospheric Pressure
Method of Line Current Isolation
Number of Output Modes
Dimensions (in.) [W x H x D]
Weight
Housing Materials and Construction
Ingress Protection
Accessories
Procedure for setting up
stimulation parameters
Number of Output Channels
Regulated Current
Microprocessor Control
User Display
Individual Output Status
Status Condition Indication
(Low Battery, Keypad, or Data)
Output Intensity Level
Output Frequency
Cycle Rate
Cycle Time
Days Remaining
TransAeris Stimulator
REF 23-1000
Synapse Biomedical, Inc.
Built-in: non-replaceable Batteries
(Stimulator disposed of after 30 days)
8
Alkaline
8 x AA 1.5v Alkaline
+5ºC to 40ºC
-20ºC to 54ºC
15% to 93%, non-condensing
700 hPa to 1060 hPa
N/A – Internally powered only.
2 – Idle or Continuous Pulse Trains
5.8" x 3.6" x 1.0"
390g
ABS
IP
55
None
TransAeris User Manual (REF 77-1000)
4
Cathodic Pulse
PEMS
LCD graphic
On, Off, Overload
Visual indicator
2% to 100% (0.1 µC to 5 µC)
6 to 20 Hz
5 to 30 Cycles per Minute (CPM)
0.5 to 1.5 sec
Count up to 30 days
(Duracell QU1500)
37
79