Release Button
Slide switch to the right
to the Lock Position
• Gently pull on the electrodes to confirm they are locked into the connector.
• If an electrode pulls out, press the release button behind the slide switch and slide
the switch back to the unlock position and repeat the process.
• Remove the adhesive backing of the surface electrode and place it on the patient in
close proximity to the percutaneous electrode exit side. (Blue = Left side of the patient,
Green = Right side of the patient)
8.3.4 Patient Cable and FrictionLoc Connector
To connect the patient cable to the FrictionLoc connector, do the following:
• Remove the TransAeris stimulator with integral patient cable from the packaging.
• Connect the color coded keyed connectors from the patient cable to the color
coded mating keyed connector from the FrictionLoc connectors.
8.3.5 Disconnecting the TransAeris System
To disconnect the cables from the TransAeris stimulator, do the following:
• Confirm that the power is OFF on the TransAeris.
• Disconnect the color coded keyed connectors from the patient cable and
FrictionLoc connectors.
• If the patient will not be using the TransAeris system anymore, dispose the TransAeris
system. DO NOT reuse.
• If the patient will be using the TransAeris system again, place the TransAeris stimulator
in a safe place where it will not be dropped or damaged.
• If the patient will be having a medical procedure that requires the FrictionLoc connectors
to be removed, press the release button and slide the switch back to the unlock position.
• Remove the surface electrodes from the patient.
• Dress the TransLoc electrodes appropriately for the medical procedure.
Figure 17 - FrictionLoc connector
FrictionLoc Hooks
67 25