12 Caution During Normal USE
12.1 CAUTION DISPLAYS
• All stimulus channels off - This display indicates when all the channels are turned
OFF suppling no stimulation.
13 Post Treatment
13.1 TRANSLOC ELECTRODE EXTRACTION
• After the conditioning treatment has been completed and prior to discharge from the
ICU, the TransLoc electrodes will be removed. Removal will be performed by the clinical
staff by pulling electrodes from the percutaneous entry point. This is done in a similar
fashion as a temporary cardiac pacing wire with a slow steady pull in a straight non
oblique line.
• To facilitate extraction, cut the TransLoc electrode at the pin and allow the coiled wires
to unravel (straighten) as you pull. This will provide a smooth path and release any
adhesions at the skin interface.
14 Cleaning
14.1 CLEANING THE TRANSAERIS STIMULATOR
Cautions:
• Clean the TransAeris stimulator when required.
• Do not immerse the TransAeris stimulator in water or cleaning agents. Severe damage
to the device may occur.
Cleaning – Follow Hospital facility procedures for all cleaning of the TransAeris.
Note: Do not expose the TransAeris to ethers, acetone, or chlorinated solvents.
These solvents may damage the case, labels, or metal components.
Figure 28 – TransAeris – All stimulus channels off