INSTRUCTIONS FOR USE
3
Equipment Setup and Use:
3.1 Prior to use:
a. Inspect the product looking for any visible damage or sharp edges that could
be caused by a drop or impact during storage.
b. Make sure product has been properly cleaned and disinfected and wiped dry
prior to each use.
3.2 Setup:
a. Position the Rail Clamps adjacent to the patient's hip.
b. Attach stirrup to the table rail by sliding the stirrup Blade Mount into the slot on
the top of the Rail Clamp. Left and Right stirrups must be in the exact same
location on opposite sides of the OR table.
NOTE: Ensure Left and Right stirrups are on the proper side; the Boots are
marked (on the bottom of the Boot) for the patient's left and right.
CAUTION: Avoid pinching fingers between rail and clamp body.
WARNING: Once the stirrup is placed on the rail, the Rail Clamp must be
tightened securely before use.
c. Pre-position the stirrup Boots by loosening the Boot Locking Mechanism and
sliding the Boot along the post until the top end (calf portion) of the Boot is
located approximately at the patient's mid-calf. Tighten the Boot Locking
Mechanism securely. Repeat above procedure for the other stirrup.
d. Position stirrup to be level with the table by twisting the Release Handle and
moving the stirrup downward.
Positioning Patients in Stirrups:
e. Ensure the patient is positioned on the table in accordance with procedure &
surgeon requirements.
Document Number: 80028261
Version: B
Blade
Rail Clamp
Page 12
Issue Date: 20 MAR 2020
Ref Blank Template: 80025118 Ver. E