CAUTION
Risk of crushing!
There is a risk of crushing to fingers when reaching into
the lung during operation.
► Only use a test unit with the test head attached.
1. Access the test software via the Start menu or the
desktop shortcut.
2. When the login screen appears, enter the user
name and password (see "Prior to first use",
page 7).
✓ After starting the program and successfully logging
on, the "Test due list" and "Test" tabs are displayed
in the function area.
5.1.3 Automatic preparation for use
Dräger recommends checking the software settings
before starting work with the test unit.
When a test is launched using the test software, the test
unit is switched on automatically for use:
– When the software is started, all sensors are zeroed.
Except for the high-pressure inlet, the connections
must be depressurized at this point.
– The artificial lung is positioned.
When the test unit is connected to the PC and
supplied with power, the status LED lights up. The
different meanings of the colors of the status LED are
explained on the help pages (F1).

5.2 During use

CAUTION
Risk of injury!
Incorrectly attached components may be ejected when
under pressure.
► Always make sure that all connections have been
correctly attached.
The test procedure is described in the test software.
Instructions for use
|
Quaestor 6000/8000

5.3 After use

5.3.1 Tasks to be done after testing
NOTICE
Do not completely close the high-pressure outlet (e.g.
with a screw plug). A pressure reducer may remain
connected.
1. Log off from the system.
2. Close the compressed air supply.
CAUTION
Risk of injury!
Injury may occur if the vent screw is not opened with
caution.
► Make sure that the vent screw is not directed
towards a person's hands or head.
3. Vent the test unit:
a. Open the vent screw on the high-pressure inlet.
b. Remove the compressed air cylinder or the
connection on the stationary compressed air
supply from the test unit.
4. Disassemble the test setup.
5.3.2 Disassembling the test unit
CAUTION
Risk of crushing!
There is a risk of crushing to fingers when reaching into
the lung during operation.
► Before dismantling the head, pull the power plug.
1. To remove the test head from the test unit, pull out
the lock and remove the test head.
2. Pull the latch of the table mount (1) backwards.
2
1
|
Use
9