Commissioning the test device
3.3.2
Setting up the test device
NOTICE
i
i
Position the test device on a table or workbench.
The PC should be positioned next to the test device so
that it is within the operator's reach.
1. Align
the
table
mount
parallel with the table edge.
The distance between the
table edge and the front
holes on the table mount
must be 12 cm.
2. Fasten the table mount to
the table using the screws
supplied.
3. Raise the test device slightly at the back.
4. Align the front feet on the test device with the slots in the
table mount and move the test device backwards in the
table mount. The back feet must engage with the table
mount.
CAUTION
!
For the test device, it is only permitted to use breathing
air in accordance with EN 12 021. If the test device is
filled with oxygen, there is a risk of fire.
5. Connect the medium pressure connection on the test
device to a medium pressure supply.
6. Connect the test device to the PC using the USB cable.
7. Connect the test device to the mains via the power supply
unit.
NOTICE
i
i
To disconnect the plug, move the socket approx. 1 mm
away from the test device. This releases the lock, and
the plug can be removed from the test device.
3.4
Preparations before every use
3.4.1
Preparing the test device
1. Remove the nebuliser from the test device and fill with
distilled water up to the fill level marking:
Remove the hose from the plug connector in a
depressurised state. To do this, slide the moveable outer
ring towards the plug connector and remove the hose.
The nebuliser can now be removed from the fastening hook.
CAUTION
!
Bacteria can develop if water is left in the nebuliser.
Only fill the nebuliser directly before testing.
12
2. When testing mask/helmet
combinations, place the
bracket (1) onto the rear of
the retainer and fix with the
hose (2).
3.4.2
i
i
12 cm
!
01521868.eps
1. Access the test software via the Start menu or the desktop
shortcut.
The startup screen appears. The program components are
loaded. If a system error is detected, this is indicated by an
error message.
A load bar with status information provides information on
the progress of the startup process.
2. When the login screen appears, enter the user name and
password.
After starting the program and successfully logging on, the
"Test due list" and "Test" tabs are displayed in the function
area.
i
i
If a test has been started and there is current being supplied to
the test device, the LED on the test device will be on weakly.
When the test device is communicating with the PC, the LED
will be on strongly. If the test device is not functional, an error
message is displayed on the PC. The LED does not come on.
3.5
The test sequence is written to the test software.
Starting the test software on the PC
NOTICE
When a test is called up on the test software, the test
device is switched on automatically. There is no on/off
switch on the test device.
CAUTION
Do not connect any respiratory protection product
to the test device before launching the software. When
starting the test device, the system must be
depressurised so that the pressure sensors can be
synchronised with the current ambient pressure.
Otherwise, no tests can be conducted. In this case,
an error message is displayed.
NOTICE
Dräger recommends checking the software settings
before starting work with the test device.
During use
1
2
00231173.eps