2.2

Product profile

The Dräger X-cal 2000 breath alcohol simulator is used to calibrate various breath alcohol tes-
ters. Using an embedded electronic device with a closed loop temperature control system, the
X-cal 2000 simulator generates breath flow to provide a precise calibration standard. The sim-
ulator also uses an alcohol reference solution, the concentration of which is very sensitive. To
accurately simulate human breath, the temperature of the water-alcohol solution used in the
solution container is maintained at 34 °C ± 0.02 °C.
2.3

Feature description

Microprocessor controlled, closed loop temperature system
7-segment Liquid Crystal Display (LCD) with a wide viewing angle
Dual temperature sensor
Maintains a precise temperature at 34 °C ± 0.02 °C
Brushless DC motor, operating at 24 VDC, capable of running 24/7
System for detecting motor failures due to mechanical and electrical defect
Audio and visual indication of system malfunction
2.4

Intended use

The X-cal 2000 simulator is used by technicians to calibrate and verify breath alcohol testing
devices. The simulator is a wet-bath calibration / verification device.
3
Use
3.1

Preparation for use

3.1.1
Alcohol reference solution
Use the solution concentration specified in the manual for breath testing devices.
Replace solution every 5 days or 20 tests.
Use of solution over time affects alcohol concentration.
Do not use a solution bottle with a broken seal, or an expired bottle.
Never use artificial methods to reheat or cool solution, or the solution container.
Keep solution at room temperature.
Do not freeze or refrigerate solution.
Do not ingest solution.
If solution is ingested, do not induce vomiting; contact your local poison control centre.
Keep solution away from eyes.
If solution comes in contact with eyes, flush eyes with water; if irritation continues, contact
your local poison control centre.
Refer to your local environmental regulations for more information about safe solution-dis-
posal amounts.
Dispose solution down a drain.