INSTRUCTIONS FOR USE
f.
Device function should be checked prior to each usage.
g. This device should only be operated by medically trained personnel.
h. The patient's toes may extend beyond the end of the stirrup boot.
i.
Use the product with specified table mount bar.
j.
Check for patient contact pressure points and consult the physician before use.
k. Use only approved pad.
l.
Secure the patient's legs in the boots prior to making adjustments.
CAUTION:
a. If it is necessary to reposition the Leg Section, one operator should always support
the patient's leg until the Leg holder is securely locked in the new position.
b. Do not exceed safe working load shown in the product specification table.
4.2 Product Specifications:
Mechanical Specifications
Product Dimensions
Material
Safe Working Load on the device
Overall Weight of Complete Device
Storage Specifications
Storage temperature
Storage Relative humidity range
Operating temperature
Operating Relative humidity range
Electrical Specifications
Not Applicable.
Software Specifications
Not Applicable.
Compatibility Specifications
The
-Matic Legholder is
Maquet
compatible with:
4.3 Sterilization Instruction:
This device is not intended to be sterilized. Equipment damage may occur.
Document Number: 80028219
Version: C
Description
16.125'' x 16.125'' x 11.625'' (41 cm x 41 cm x
29.5 cm)
Stainless Steel, Alloy Steel, Aluminum, Engineering
polymers, Bronze, Low Density Polyethylene, Velcro
350 lbs (159 kg)
16.5 lbs (7.5 kg) each
Description
-29ᵒ C to +60ᵒ C
15% to 85%
This device is intended to be used in a controlled
Operating Room environment.
Description
Not Applicable.
Description
Not Applicable.
Description
Square table mount bar with 1.25" x 1.25"
(3.2 cm x 3.2 cm) cross section
Page 15
Issue Date: 10 AUG 2021
Ref Blank Template: 80025118 Ver. E