Message
70-129 mg/dL
(3.9-7.2 mmol/L)
130-239 mg/dL
(7.2-13.3 mmol/L)
>= 240 mg/dL (13.3 mmol/L)
> 600 mg/dL (33.3 mmol/L)

Error Message

Error Message Cause
The batteries cannot
provide enough power
for a test.
Strip has been used.
Problem in operation.
What it Means
General
70-179 mg/dL
70-119 mg/dL
(3.9-9.9 mmol/L)
(3.9-6.6 mmol/L)
General
180-239 mg/dL
120-239 mg/dL
(9.9-13.3 mmol/L)
(6.7-13.3 mmol/L)
What To Do
Recharge the battery
immediately and reset
date and time on the
meter setting.
Repeat the test with a
new strip.
Review the instructions
and repeat the test with
a new strip. If problem
persists, contact the local
customer service for
assistance.
You may have removed
the strip after applying
blood, or insufficient
blood volume.
Ambient temperature
is below the system's
operation range.
Ambient temperature
is above the system's
operation range.
Announcement of Error Message by Universal
Tone
• Hi or Lo Result: Results larger than 600 mg/dL (33.3 mmol/L) are
represented as 999, i.e. three groups of nine short beeps with
pauses between the groups.
• Results lower than 20 mg/dL (1.1 mmol/L) are represented as
000, i.e. three long beeps.
• Low Battery Warning: When the batteries are about to be
exhausted, 2 quick beeps are announced three times in
succession. This warning is sounded when the meter is switched
on.
• Other Errors: Other Error Messages are announced by 2 quick
beeps four times in succession.
EN-13
Review the instructions
and repeat test with a
new test strip.
System operational range
is 10°C to 40°C (50°F to
104°F). Repeat the test
after the device and test
strip have reached the
above temperature.