Device Storage
• Storage condition: -20°C to 60°C (-4°F to 140°F), below 95%
relative humidity.
• Always store or transport the device in its original storage
case.
• Avoid dropping and heavy impact.
• Avoid direct sunlight and high humidity.
Meter Disposal
The used meter should be treated as contaminated and may
carry a risk of infection during measurement. The batteries in
this used meter should be removed and the meter should be
disposed in accordance with local regulations.
The meter falls outside the scope of the European Directive
2002/96/ EC-Directive on waste electrical and electronic
equipment (WEEE).
Caring for Your Test Strips
• Storage condition: 2°C to 32°C (35.6°F to 89.6°F) for glucose
test strip, below 85% relative humidity. Do NOT freeze.
• Store your test strips in their original vial only. Do not transfer
to another container.
• Store test strip packages in a cool and dry place. Keep away
from direct sunlight and heat.
• After removing a test strip from the vial, immediately close the
vial cap tightly.
• Touch the test strip with clean and dry hands.
• Use each test strip immediately after removing it from the vial.
• Write the opening date on the strip vial label when you first
open it. Discard remaining test strips after 6 months.
• Do not use test strips beyond the expiry date. This may cause
inaccurate results.
• Do not bend, cut, or alter a test strip in any way.
• Keep the strip vial away from children since the cap and the
test strip may be a choking hazard. If swallowed, promptly see a
doctor for assistance.
For further information, please refer to the test strip package
insert.
Important Control Solution Information
• Use only our control solutions with your device.
• Do not use the control solution beyond the expiry date or
3 months after first opening. Write the opening date on the
control solution vial and discard the remaining solution after 3
months.
• It is recommended that the control solution test be done at
room temperature 20°C to 25°C (68°F to 77°F). Make sure your
control solution, device, and test strips are at this specified
temperature range before testing.
• Shake the vial before use, discard the first drop of control
solution, and wipe off the dispenser tip to ensure a pure sample
and an accurate result.
• Store the control solution tightly closed at temperatures
between 2°C to 30°C (35.6°F to 86°F). Do NOT freeze.
EN-11