Error Message

Error Message Cause
The batteries cannot
provide enough
power for a test.
Strip has been used.
Problem in
operation.
You may have
removed the strip
after applying blood,
or insufficient blood
volume.
Ambient
temperature is
below the system's
operation range.
Ambient
temperature is
above the system's
operation range.
What To Do
Recharge the battery
immediately and reset
date and time on the
meter setting.
Repeat the test with a
new strip.
Review the
instructions and
repeat the test with a
new strip. If problem
persists, contact the
local customer service
for assistance.
Review the
instructions and
repeat test with a new
test strip.
System operational
range is 10°C to
40°C (50°F to 104°F).
Repeat the test after
the device and test
strip have reached the
above temperature.
EN-16
Announcement of Error Message by
Universal Tone
• Hi or Lo Result: Results larger than 600 mg/dL (33.3
mmol/L) are represented as 999, i.e. three groups of nine
short beeps with pauses between the groups.
• Results lower than 20 mg/dL (1.1 mmol/L) are represented
as 000, i.e. three long beeps.
• Low Battery Warning: When the batteries are about to be
exhausted, 2 quick beeps are announced three times in
succession. This warning is sounded when the meter is
switched on.
• Other Errors: Other Error Messages are announced by 2
quick beeps four times in succession.