Result Reading (for β-Ketone test)
Message
< 0.1 mmol/L
0.1 to 8.0 mmol/L
> 8.0 mmol/L

Error Message

Error Message Cause
The batteries cannot
provide enough power
for a test.
Strip has been used.
Problem in operation.
You may have removed
the strip after applying
blood, or insufficient
blood volume.
HCT>75%
What it Means
What To Do
Replace the batteries
immediately.
Repeat the test with a
new strip.
Review the instructions
and repeat the test with
a new strip. If problem
persists, contact the local
customer service for
assistance.
Review the instructions
and repeat test with a
new test strip.
Wrong code strip
is inserted or other
coding errors.
Expired code strip.
Ambient temperature
is below the system's
operation range.
Ambient temperature
is above the system's
operation range.
EN-14
Make sure to insert the
right code strip for the
meter. Check if the code
numbers on the code strip
and the strip vial label are
the same.
Repeat the test with a
new code strip.
System operational range
is 10°C to 40°C (50°F to
104°F). Repeat the test
after the device and test
strip have reached the
above temperature.