RESULTS stored in the memory. Failure to do so will mix up the
blood glucose test results with the control solution test results
in the memory.
3. Apply the control solution. Shake the control solution vial
thoroughly before use. Squeeze out a drop and wipe it off,
then squeeze another drop and place it on the tip of the vial
cap. Hold the device to move the absorbent hole of the test
strip to touch the drop. Once the confirmation window is filled
completely, the device will begin counting down. (g)
Note:
To avoid contaminating the control solution, do not directly
apply the control solution onto a strip.
4. Read and compare the result. After counting down to 0, the
test result of the control solution will appear on the display.
Compare this result with the range printed on the test strip vial
or individual foil pack and it should fall within this range. If the
test result is out of range, read the instructions again and repeat
the control solution test. (h)
Note:
• Do NOT test your blood.
• The control solution range printed on the test strip vial or
individual foil pack is for control solution use only. It is not a
recommended range for your blood glucose level.
• Refer to the Maintenance section for important information
about your control solutions.
Out-of-range results:
If you continue to get results that fall outside the range printed on
the test strip vial, it means that the meter and strips might not be
working properly.

Testing Your Blood Glucose

Test Strip Appearance
Absorbent Hole
Apply a drop of blood
here. The blood will be
automatically absorbed.
Test Strip Handle
Hold this part to insert
the test strip into the slot.
Inserting a Test Strip
Insert the test strip into its slot.
Important!
The front side of test strip should face up when inserting the test
strip. Test results might be wrong if the contact bar is not fully
inserted into the test slot.
EN-5
Confirmation Window
This is where you confirm
if enough blood has been
applied to the absorbent hole
in the strip.
Contact Bars
Insert this end of the test strip
into the meter. Push it in firmly
until it will go no further.