without ketosis. Please consult the healthcare professional
before use.
• The measurement unit used for indicating the concentration
of blood glucose can have mg/dL or mmol/L. The approximate
calculation rule for conversion of mg/dL in mmol/L is:
mg/dL
Divided by 18
mmol/L
Times 18
For example:
1) 120 mg/dL ÷ 18 = 6.6 mmol/L
2) 7.2 mmol/L x 18 = 129 mg/dL approximately
Introduction
Intended Use
This system is intended for use outside the body (in vitro
diagnostic use) by people with diabetes at home and by health-
care professionals in clinical settings as an aid to monitor the
effectiveness of diabetes control. It is intended to be used for the
quantitative measurement of glucose (sugar) in fresh capillary
(from the finger, palm, forearm, and upper arm) and venous blood
samples. It should not be used for the diagnosis of diabetes, or
testing on newborns.
Professionals may test with capillary and venous blood sample.
Use only heparin for anticoagulation of whole blood.
Home use is limited to capillary blood from the finger tip.
= mmol/L
= mg/dL
Test Principle
Your system measures the amount of sugar (glucose) in whole
blood. The glucose testing is based on the measurement of
electrical current generated by the reaction of glucose with the
reagent of the strip. The meter measures the current, calculates
the blood glucose level, and displays the result. The strength of
the current produced by the reaction depends on the amount of
glucose in the blood sample.
Product Overview
3. Display Screen
5. Test Strip Slot
EN-2
1. SET Button 2. USB Port
4. MAIN Button
6. Bluetooth Indicator
(for DM30b)