P P r r o o b b l l e e m m
Patient temperature
Patient temperature output
(PTO)
8001-109-005 Rev AA.3
P P o o s s s s i i b b l l e e c c a a u u s s e e
Out of range
External device output
displays high value >45° C
when input is out of range
(As a result of one of the
following: patient probe
disconnected, controller in
standby / sleep mode,
patient temperature outside
the range of 25° - 45°C).
A A c c t t i i o o n n
Check the placement of the
probe
To resume calibration:
Disconnect the external
device from the reusable
adapter cable
Tap the Help button to
display the alarm
screens
Find the "Temperature
Output Alarm" screen
Tap Confirm to restart
calibration
Wait until the "Monitor" is
solid
Connect the external
device to the reusable
adapter cable
R R e e m m o o v v e e f f r r o o m m u u s s e e
Not applicable
Not applicable
51
EN