Hose connectors
Power cord
Hose and power cord management straps
Storage compartment door
Inside storage compartment
Graphical user interface display
Handle
14. Follow specified contact time in accordance with the disinfectants manufacturer's instructions for use.
15. To rinse, wipe the hoses and patient temperature probe cables, with a soft, lint free cloth moistened with sterile distilled
water.
16. To rinse, wipe the controller, reservoir and reservoir lid surfaces, and system components with a lint free cloth
moistened with sterile distilled water.
17. To dry, wipe the controller, reservoir, reservoir lid surfaces, and system components with a clean, dry cloth to remove
any excess liquid.
18. Replace the reservoir.
19. Allow the external surfaces of the controller and components to dry thoroughly.
20. Store the power cord, cables, and hoses.
EN
42
8001-109-005 Rev AA.3