Troubleshooting; Technical Specifications - Renfert Waxlectric I Traduction Du Mode D'emploi D'origine

Masquer les pouces Voir aussi pour Waxlectric I:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 30
7

Troubleshooting

Trouble
EN
Display does not go
on, unit does not
operate.
Unit operating, but
no temperature at
the tip
The tip fails to
reach the desired
temperature.
Values are not
saved.
Pressing the P key
always displays
the same tempera-
ture.
8

Technical Specifications

Supply voltage:
Power consumption:
Min. operating temperature:
Max. operating temperature:
Transformer:
- Input voltage:
- Output voltage:
Dimensions (W x H x D):
Weight, approx.:
- with Transformer:
- without Transformer:
Possible cause
• Control unit not switched on�
• Power supply interrupted�
• The plug is not fully inserted into the control
unit� �
• Faulty control unit�
• Faulty transformer�
• Short circuit in the tip�
• Tip not properly screwed in�
• Tip / handpiece contact contaminated�
• Faulty tip�
• Handpiece cable not properly plugged into
the control unit�
• Plug on handpiece cable shows signs of
corrosion�
• Incorrect power supply to the tip because the
OEM mains power supply unit is not being
used�
• Memory chip faulty�
• The same temperature has been saved to
each memory location�
• Switch the control unit on at key A�
• Check the power supply and reconnect, as required�
• Insert the plug until it clearly snaps into position�
• Have unit replaced by Service Department�
• Replace with spare (refer to the spare parts list)�
• Clean the tip and handpiece (use compressed air)�
• Replace the tip�
• Faulty heater handle --> Replace with spare (refer to
the spare parts list)�
• Hand tighten tip�
• Clean the contact (use compressed air)�
• Replace the O-ring (refer to the spare parts list)�
• Replace with a new tip (refer to sec� 3�5 Accessories)�
• Fully insert the handpiece cable into the socket until
the cable locks in place�
• Reinsert the plug, clean the contacts, if required�
• Use the OEM mains power supply unit�
• Have the unit repaired�
• Assign different temperatures to each memory loca-
tion (refer to Sec� 5�5)�
Waxlectric I
5 W
50 °C [122 °F]
200 °C [392 °F]
100 - 240 VAC, 50 / 60 Hz
12 VDC / 1 A
120 x 60 x 98 mm
[4�72 x 2�36 x 3�86 inch]
430 g / [15�2 oz]
240 g / [8�5 oz]
- 10 -
Corrective action
Waxlectric II
12 VDC
10 W
150 x 60 x 98 mm
[5�91 x 2�36 x 3�86 inch]
460 g / [16�2 oz]
270 g / [9�5 oz]

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Waxlectric ii

Table des Matières