70
Voltage output
Load
Ripple
Monitor Output
Uninterrupted monitoring of the output current (e.g. using a multimeter) or connection of a
floating local indicator.
An open circuit in the monitor output (e.g., due to a line break) has no impact on the current
output.
Max. burden
Maximum permissible
cable length
Transmission Behavior
Gain error
Gain error
due to temperature
Offset voltage
(voltage output)
Offset current
(current output)
Offset drift
due to temperature
±10 V
Linear up to ±12 V Max. ±15 V
0 ... 10 V
Linear up to 12 V
±5 V
Linear up to ±6.5 V Max. ±15 V
0 ... 5 V
Linear up to 6.5 V
Min. 1 kΩ
Short-circuit-proof Short-circuit current < 70 mA
2.5 mV
rms
10 Ω
Max.
voltage drop
3 m
≤ 0.1 %
of meas. value
≤ 50 ppm/K
of meas. value
≤ 10 mV
≤ 20 µA
≤ 50 ppm/K
of full scale output Reference temperature 23 °C
Specifications
Max.
15 V
Max.
15 V
0.3 V
at 23 °C
Reference temperature 23 °C
at 23 °C
at 23 °C