Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007 3rd Edition, Table 4)
Electromagnetic Immunity
Clario is intended for use in the electromagnetic environment specified below� The cus-
tomer or the user of Clario should assure that Clario is used in such environment�
Immunity
IEC 60601
Tests
test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2�5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies�
NOTE 2 These guidelines may not apply in all situations� Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people�
Field strengths from fixed RF transmitters, such as base stations for radio (cellular / cordless)
a
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy� To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered� If the measured
field strength in the location in which Clario are used exceeds the applicable RF compliance level
above, Clario should be observed to verify normal operation� If abnormal operation is observed,
additional measures may be necessary, such as reorienting or relocating Clario�
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m�
b
328 |
Technical documentation
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications
equipment should be used no closer to any
part of Clario, including cables, than the rec-
ommended separation distance calculated from
the equation applicable to the frequency of the
transmitter�
Recommended separation distance
3 Vrms
d = 1.2 √ P
d = 1.2 √ P
80 MHz to 800 MHz
d = 2.3 √ P
3 V/m
800 MHz to 2�5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in meters (m)
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site surveya,
should be less than the compliance level in
each frequency range�
Interference may occur in the vicinity of equip-
ment marked with the following symbol:
b